Lynx E

A solution for rapid X-ray emission spectroscopy

This device uses a Johann-type spectrometer with cylindrically bent analyser crystals. The setup is made dispersive by positioning the sample inside the Rowland-circle. Note that this solution allows us to make use of the X-ray photons more effectively.

Lynx E Specification

Energy Range

4.5 to 13 keV

Measurement

XES in an off-Rowland arrangement

X-ray Source

2 kW water-cooled diffraction tube

Beam Size at Sample

⌀ 6 mm

Analyzer Crystals

cylindrically curved Johann crystals, with a radius of 250 mm

Detector

Spatially resolving 2D detector

Beam Path

air

Footprint and Height

1.2 m ×  1.2 m; 1.8 m

Weight

550 kg

Lynx E Features

– detailed information about the local atomic structure around your element of choice
– high spin and valence sensitivity with element selectivity
– simple, scanning-free XES measurements on bulk samples
– easily adjustable to different elements and emission features
– energy range between 5 and 13 keV (Kα, Kβ, K-vtc: Ti – Br)
– the chemically most sensitive valence-to-core emission features can also be recorded
– compatible with various possible sample environments
– automatic sample exchanger 